Water Hardness
|
Measuring range
Resolution
Accuracy
Temp. compensation
Calibration Solution
Calibration Mode
|
(0 ~ 10)mmol/L; (0 ~ 401)mg/L(Ca);
(0 ~ 1000)mg/L(CaCO3); (0 ~ 100)ºfH(France Degree);
(0 ~ 561)mg/L(CaO); (0 ~ 56)ºdH(German Degree);
(0 ~ 20)mmol/L(Boiler); (0 ~ 70)ºeH (England Degree)。
0.01 and 0.1 water hardness units
±5% FS
(5 ~ 50)℃((automatic))
B1 Calibration Solution — 2.00×10-2 mmol/L
B2 Calibration Solution — 2.00×10-1 mmol/L
B3 Calibration Solution — 2.00 mmol/L
(B3Calibration Solution also can be used for the soaking solution of electrode)
(a)B1/B2 Calibration—using B1 and B2 Calibration Solution,suitable for < 2.00×10-2 mmol/L low concentration water quality,for example, boiler
water.
(b)Calibration—using B2 and B3 Calibration Solution,suitable forgeneral water quality.
|
Temperature
|
Measuring range
Resolution
Accuracy
|
0 ~ 60℃
0.1℃
±0.5℃
|
Other Technical Parameter
|
Data storage
Content Storage
Power Dimension and weight Quality & safecertificate |
99 groups
serial number of measuring value, measuring value, units, temperature, ATC or MTC, date, time and calibration way. Two AA cells(1.5V×2) Meter:(65×120×31)mm/180g; Carry Case: (255×210×50)mm/790g IISO9001:2000 and CE |
working conditions
|
Ambient temperature
ambient humidity
|
(5~35)℃
≤85%RH
|
1. Model YD300 water hardness meter |
1set |
2.Model 601-S water hardness electrode( With two measruing heads and one screw plug ) |
1pc |
3. AA cell (1.5V ) |
2pcs (included) |
4.100mL hardness calibration solution ( B1, B2 and B3 ) |
1btl/each |
5.20mL test cup |
6pcs |
6. Test cup pedestal |
1pc |
7.Carring case |
1pc |
8. Operation manual |
1 set |